Relationship between end-tidal carbon dioxide and arterial... : Medicine (2024)

1 Introduction

End-expiratory carbon dioxide partial pressure (PETCO2) monitoring has been widely used in intensive care units (ICU) for patients in need of mechanical ventilation. In healthy people, the correlation between PETCO2 and arterial carbon dioxide tension (PaCO2) is strong. In sick patients, however, the variation is large, which may be associated with many factors.[1–3] PETCO2 can exceed PaCO2 or can be lower than PaCO2.[4] When ventilator settings are adjusted, the correlation can be very poor or even reversed in ventilated patients. The ventilation of emergency patients can only be adjusted according to values derived from an arterial blood gas analysis. PETCO2 measurements cannot be regarded as absolutely accurate values, except maybe in patients without primary cardiorespiratory dysfunction.[1]

So far, the relationship between PaCO2 and PETCO2 has been reported in several pre-clinical[5–7] and clinical studies, including children/infants,[8–11] or mechanically ventilated patients with a single disease (acute respiratory distress syndrome, neurosurgical, or postcardiac arrest patients).[12–15] Only 1 large sample study[16] analyzed the relationship between PaCO2 and PETCO2 in 219 arterial blood gases obtained from 87 patients. They found a good correlation between the mean of PETCO2 and PaCO2 when using synchronized intermittent mandatory ventilation (SIMV), continuous positive airway pressure, and T-Tube models; SIMV (r = 0.893, 0.841, and 0.923, respectively).[16] However, they did not conduct stratified analysis according to the disease type and severity, and its clinical significance was limited. Thus, in this study, we further examined the correlation between PaCO2 and PETCO2 under different ventilator modes, different disease types, and different oxygenation indexes in mechanically ventilated patients with relatively stable conditions in ICU.

2 Materials and methods

2.1 Study design

This was a cross-sectional study that evaluated the patients on invasive mechanical ventilation admitted to ICU of a tertiary university hospital between June 2018 and March 2019.

2.2 Eligibility criteria

Inclusion criteria were as follows: all patients underwent endotracheal intubation or tracheotomy; continued application of ventilator ≥48 hours in ICU; no vasoactive drugs were prescribed; age >18 years.

Exclusion criteria: incomplete data or patient refusal to participate.

The Institutional Review Board of Harrison International Peace Hospital (2018-1-013) approved the study protocol. Informed consent was obtained from all individual participants included in the study.

2.3 Interventions and data collection

Patients received one of the following mechanical ventilation methods: assisted/controlled ventilation (A/C), SIMV, and spontaneous breathing (SPONT) mode. The mainstream PETCO2 monitor (KMI605A, Beijing Jinjiaxing Co., Ltd., Beijing, China) was used by the same doctor (with 10 years of experience) to detect PETCO2. The sampling sensor was directly connected to the Y-shaped pipe of the ventilator and the endotracheal intubation or tracheotomy catheter. Blood samples of radial artery or femoral artery were collected and the arterial blood gas analysis was completed at beside (ABL90, Leidu, Denmark). PETCO2 and arterial blood gas analysis were completed within 5 minutes.

2.4 Outcomes

Primary outcome measures: PETCO2 and PaCO2 were collected as primary outcome variables. The difference and correlation between PETCO2 and PaCO2 were tested in all patients.

Secondary outcome measures: The following data were collected when PETCO2 and PaCO2 were recorded, including age, gender, body mass index, primary disease type, mean arterial pressure, heart rate, ventilator mode and oxygenation index, Charlson comorbidity index, acute physiology and chronic health evaluation scoring system II, sequential organ failure assessment and treatments.

2.5 Statistical analysis

Assuming a type 1 error of 5% (alpha of 0.05), a power of 90% and r = 0.50 in preliminary, this study would require a sample size of 40 patients. To account for dropouts and incomplete data, we aimed for a sample size of 100 patients. Sample size was calculated based on both primary outcomes and the larger of the 2 calculations was utilized.

The Shapiro–Wilk test was used to verify whether all recorded variables were normally distributed (P > .05). Continuous data are expressed as the mean ± standard deviation. Correlations among data with measurable outcomes were analyzed using the Pearson test if distributed normally, or as median (interquartile range) and with the Spearman test if non-normally distributed. When there was a quantitative relationship between the 2 variables, linear regression was used to explore the regression equation. A P value <.05 was considered statistically significant.

3 Results

3.1 Patients

A total of 184 patients with 298 pairs of PETCO2-PaCO2 data were included in the analysis. The mean age was 68.45 ± 16.50 years, and 126 patients (68.48%) were men. Main characteristics of the patients are shown in Table 1.

Table 1 - Main characteristics of 184 subjects.

Variables Value
Age (y) 68.45 ± 16.50
Gender, male, no. (%) 126 (68.48)
BMI (kg/m2) 23.43 ± 3.35
Temperature, °C 36.58 ± 1.67
Mean arterial pressure, mm Hg 62.66 ± 7.34
Heart rate, beats/min 90.82 ± 4.24
Charlson comorbidity index 2.12 ± 1.31
SOFA score 7.67 ± 2.20
APACHE II score 14.58 ± 3.44
Vasopressor, no. (%) 45 (24.46)
CRRT, no. (%) 31 (16.85)

APACHE II = acute physiology and chronic health evaluation scoring system II, BMI = body mass index, CRRT = continuous renal replacement therapy, SOFA = sequential organ failure assessment.


3.2 Correlation analysis of PETCO2 and PaCO2 under different ventilator modes

Without distinguishing the ventilator mode, there was a significant positive correlation between PETCO2 and PaCO2; the correlation coefficient was 0.72, the linear regression equation Y = 11.81 + 0.65x (Y: PETCO2; x: PaCO2) (Table 2 and Fig. 1). As shown in Figure 1, the majority of PETO2 and PaCO2 values are distributed in 20 to 60 mm Hg (black dotted frame).

Table 2 - Correlation analysis of PETCO2 and PaCO2 under different ventilator modes.

Ventilator modes Pairs PETCO2 (mm Hg) PaCO2 (mm Hg) PaCO2- PETCO2 gap (mm Hg) r P
All modes 298 36.65 ± 10.31 39.40 ± 11.30 1.60 ± 7.60 0.72 <.001
A/C 30 37.33 ± 7.85 38.56 ± 8.01 1.23 ± 8.16 0.47 <.001
SIMV 127 36.20 ± 10.63 38.06 ± 12.76 1.86 ± 7.42 0.81 <.001
SPONT 68 38.16 ± 8.58 39.56 ± 7.22 1.40 ± 7.57 0.55 <.001

A/C = assisted/controlled ventilation, PaCO2 = arterial carbon dioxide tension, PETCO2 = end-expiratory carbon dioxide partial pressure, SIMV = synchronized intermittent mandatory ventilation, SPONT = spontaneous breathing.


When comparing different ventilator modes, only the SIMV mode showed a significant correlation (r = 0.81, P < .001). In both A/C and SPONT mode, the correlation was relatively weak (correlation coefficient r = 0.47 and 0.55, respectively).

3.3 Correlation analysis of PETCO2 and PaCO2 of different disease types

In patients with chronic obstructive pulmonary disease, multiple injuries, severe pneumonia, gastrointestinal surgery, and cerebrovascular diseases, PETCO2 and PaCO2 were positively correlated (the correlation coefficients were 0.80, 0.64, 0.60, 0.57, and 0.53 respectively; Table 3). For other diseases (including malignant tumors and cardiovascular disease), no correlation was found (r = 0.46, P = .06).

Disease types Cases PETCO2 (mm Hg) PaCO2 (mm Hg) PaCO2- PETCO2 gap (mm Hg) r P
COPD 70 42.81 ± 12.04 48.16 ± 13.14 5.35 ± 7.97 0.80 <.001
Multiple trauma 23 41.30 ± 8.14 38.75 ± 5.50 –2.55 ± 6.25 0.64 .001
Severe pneumonia 55 32.05 ± 7.39 35.13 ± 8.13 3.08 ± 6.97 0.60 <.001
Gastrointestinal surgery 73 36.77 ± 7.53 36.13 ± 7.22 –0.64 ± 6.82 0.57 <.001
Cerebrovascular disease 56 34.79 ± 7.37 35.01 ± 7.85 0.22 ± 7.36 0.53 <.001
Others 21 32.52 ± 6.65 33.71 ± 6.77 1.19 ± 6.95 0.46 .06

COPD = chronic obstructive pulmonary disease, PaCO2 = arterial carbon dioxide tension, PETCO2 = end-expiratory carbon dioxide partial pressure.


3.4 Correlation analysis of PETCO2 and PaCO2 with different oxygenation indexes

Oxygenation index <200 mm Hg, correlation coefficient r = 0.69, P < .001; oxygenation index ≥200 correlation coefficient r = 0.73, P < .001 (Table 4). Under different oxygenation indexes, there was no statistically significant difference between the 2 correlation coefficients (Z = 0.67, P = .50). Among 116 pairs of data with oxygenation index <200 mm Hg, the difference of PaCO2-PETCO2 ≥10 mm Hg was found in 25 pairs (21.55%); in 182 pairs of data with oxygenation index ≥200 mm Hg, the difference of PaCO2-PETCO2 ≥10 mm Hg was found in 26 pairs (14.29%) (χ2 = 2.64, P = .19).

Table 4 - Correlation analysis of PETCO2 and PaCO2 with different oxygenation indexes.

Oxygenation indexes (mm Hg) Cases PETCO2 (mm Hg) PaCO2 (mm Hg) PaCO2- PETCO2 gap (mm Hg) r P
<200 116 38.72 ± 10.23 40.86 ± 10.47 2.14 ± 8.09 0.69 <.001
≥200 182 35.90 ± 8.98 37.15 ± 10.34 1.25 ± 7.26 0.73 <.001
Z = 0.67, P = .50

Comparison of correlation coefficients under different oxygenation indexes.PaCO2 = arterial carbon dioxide tension, PETCO2 = end-expiratory carbon dioxide partial pressure.


4 Discussion

This study showed that there was a significant positive correlation between PETCO2 and PaCO2 on invasive mechanical ventilation admitted to ICU, especially in SIMV mode, chronic obstructive pulmonary disease patients. Under different oxygenation indexes, the correlation remained strong.

In healthy people, the difference between PETCO2 and PaCO2 is generally 2 to 5 mm Hg.[17] PaCO2, dead space, lung perfusion, and sampling points affect PETCO2. When the dead space is large, the alveolar CO2 (PACO2) evacuation is not uniform, so PETCO2 is more likely to have a lower value compared with PaCO2. When the ventilation/perfusion ratio is low, the effect of shunt on PETCO2 is small, and cannot easily degrade PETCO2. The difference between PETCO2 and PaCO2 in patients with respiratory failure is large, and the difference is closely related to ventilation/perfusion. PETCO2 should not be used to evaluate PaCO2. In severely ill patients, pulmonary organic diseases cause increased pulmonary shunts; this mixed blood flow into the arterial system results in an increased gradient of PETCO2-PaCO2 difference.[4] In severe lung diseases or systemic diseases, this difference is as high as 20 mm Hg. In other words, PETCO2 underestimates PaCO2 levels. Sivan et al[18] found that the average difference is 3.4 ± 6.6 mm Hg. When the PaCO2/PACO2 ratio is lower than 0.3, the difference begins to increase, reaching 7.8 ± 7.3 mm Hg; when the PaCO2/PACO2 ratio is greater than 0.3, the difference is only 0 ± 3.4 mm Hg. This study did not distinguish ventilator mode and disease type, the difference was 2.75 ± 8.38 mm Hg, but stratified analysis was not based on PaCO2/PACO2.

4.1 Ventilator mode and PETCO2

Weinger and Brimm[19] found a good correlation between PaCO2 and PETCO2 in 25 adult patients with lung disease or extrapulmonary disease using a SIMV mode; the difference between PaCO2 and PETCO2 was 4.24 ± 4.42 mm Hg. In patients with non-pulmonary diseases, who underwent mechanical ventilation or automatic ventilation through tracheal intubation the difference between PaCO2 and PETCO2 was 0.8 to 3.5 mm Hg.[20] A recent study suggested a strong correlation between PaCO2 and PETCO2 under the conditions of SIMV, continuous positive airway pressure mode, and T-tube,[16] which was consistent with our findings. However, there was only a weak correlation between PaCO2 and PETCO2 in A/C mode. This might be due to small sample size; only 30 sets with A/C mode were analyzed. In addition, the A/C mode in this study was mostly used for surgical postoperative or severe pneumonia patients, which requires deep sedation, analgesia, and complete control of ventilation, and alveolar minute ventilation and exhaled tidal volume are basically in a constant state, so PETCO2 variation is small in this case. Even if PaCO2 increases or decreases, it is difficult to stimulate the central or surrounding receptors, change the breathing frequency and tidal volume, so there is weak correlation between PaCO2 and PETCO2.

4.2 Disease type and PETCO2

Kerr et al[21] reported a good correlation between PaCO2 and PETCO2 in adult patients with traumatic brain injury without pulmonary disease (positive end-expiratory pressure <5 cmH2O). Another study found that PaCO2 and PETCO2 had a strong correlation regardless of the disease, using a ventilator, or SPONT; but the correlation coefficients were different among different diseases.[22] Barton et al reported that in non-intubated patients with different conditions in the emergency room, PaCO2 and PETCO2 also had a strong correlation. PETCO2 monitoring may be sufficient to represent PaCO2 and avoid repeated arterial blood gas analysis.[22] Tobias and Meyer[23] found that percutaneous CO2 monitoring is more accurate than PETCO2 in predicting PaCO2 in infants and young children. The difference between percutaneous CO2 and PaCO2 is smaller than the deviation between PaCO2 and PETCO2 (2.3 ± 1.3 mm Hg and 6.8 ± 5.1 mm Hg, respectively). Continuous monitoring of PETCO2 and finger oxygen saturation is safe and effective for patients after coronary artery bypass grafting. Moreover, PETCO2 can predict PaCO2 (r = 0.76), can easily detect hypercapnia, and has a sensitivity of 95%.[24] Consistently, in this study, we found a good correlation between PaCO2 and PETCO2 in patients with COPD, multiple injuries, and severe pneumonia. A correlation coefficient of 0.57 and 0.53 for gastrointestinal surgery and cerebrovascular diagnosis is weak, and clinically irrelevant, even though statistically significant.

4.3 Oxygenation index and PETCO2

Previous studies have argued that the relationship between PETCO2 and PaCO2 in different clinical settings is controversial. McDonald et al[25] suggested a good correlation between PETCO2 and PaCO2 in 129 critically ill patients who received invasive mechanical ventilation through tracheal intubation. The statistical analysis of 1708 paired data showed a higher PETCO2 (39.9 ± 12.7 mm Hg) compared to PaCO2 (45.5 ± 14.1 mm Hg); PETCO2-PaCO2 difference was ≤5 mm Hg in 54%, and ≤10 mm Hg in 80% paired data. The presence of lung disease had a negative impact on the correlation between the two. In the data of 640 groups with oxygenation index <200 mm Hg, the difference of PETCO2-PaCO2 in 223 groups (35%) was >10 mm Hg. However, among the 1068 data sets with an oxygenation index >200 mm Hg, only 111 groups (10%) had a difference >10 mm Hg. This trend suggests that the lower the oxygenation index, the greater the difference between the two.[25] In this study, among 116 pairs of data with oxygenation index <200 mm Hg, the difference of PaCO2-PETCO2 ≥10 mm Hg was found in 25 pairs (21.55%); in 182 pairs of data with oxygenation index ≥200 mm Hg, the difference of PaCO2-PETCO2 ≥10 mm Hg was found in 26 pairs. These data suggest that the oxygenation index of adult patients was negatively correlated with the PaCO2-PETCO2 difference.

This study had several limitations. First of all, the PETCO2 sampling sensor was directly connected to the Y-shaped pipe of the ventilator. It is necessary to ensure that the exhaled gas will not leak due to insufficient tube cuff pressure, which may be ignored by the researcher during the research process, resulting in measurement error. However, in all patients during mechanical ventilation, manually re-measure the cuff pressure every 6 to 8 hours, and the pressure is always maintained at 25 to 30 cmH2O to minimize the possibility of air leakage. Second, although PETCO2 was measured only once without the average of multiple measurements, but when recording the value, ensure that the PETCO2 is in a steady state (the fluctuation range is <±2 mm Hg within 5 minutes), which may reduce the error. Finally, with the increasing accuracy of the measuring instruments, PETCO2 has become clinically applied as a substitute for PaCO2. However, caution is required for its application, and use without knowing the advantages and disadvantages of this method may result in erroneous results and improper clinical interpretation. Further studies are needed to assess their suitability in different diseases and clinical situations.

5 Conclusions

In patients receiving invasive mechanical ventilation, PETCO2 and PaCO2 showed a good correlation in different ventilator modes, different disease types and different oxygenation indexes, especially in SIMV mode and chronic obstructive pulmonary disease patients.

Author contributions

Conceptualization: Jinrong Wang, Jianjun Zhang, Zhaobo Cui.

Data curation: Jianjun Zhang, Yajing Liu, Huimian Shang, Li Peng, Zhaobo Cui.

Formal analysis: Jinrong Wang, Jianjun Zhang, Yajing Liu, Huimian Shang, Li Peng, Zhaobo Cui.

Funding acquisition: Jinrong Wang, Zhaobo Cui.

Investigation: Jianjun Zhang, Zhaobo Cui.

Methodology: Jianjun Zhang, Zhaobo Cui.

Project administration: Jinrong Wang, Zhaobo Cui.

Writing – original draft: Jianjun Zhang, Zhaobo Cui.

Writing – review & editing: Jinrong Wang, Yajing Liu, Huimian Shang, Li Peng.

References

[1]. Prause G, Hetz H, Lauda P, Pojer H, Smolle-Juettner F, Smolle J. A comparison of the end-tidal-CO2 documented by capnometry and the arterial pCO2 in emergency patients 1997;35:145–8.

[2]. Kartal M, Goksu E, Eray O, et al. The value of ETCO2 measurement for COPD patients in the emergency department. Eur J Emerg Med 2011;18:09–12.

[3]. Lee SW, Hong Y-S, Han C, et al. Concordance of end-tidal carbon dioxide and arterial carbon dioxide in severe traumatic brain injury. J Trauma 2009;67:526–30.

[4]. Yamanaka MK, Sue DY. Comparison of arterial-end-tidal PCO2 difference and dead space/tidal volume ratio in respiratory failure. Chest 1987;92:832–5.

[5]. Tusman G, Suarez-Sipmann F, Bohm SH, Borges JB, Hedenstierna G. Capnography reflects ventilation/perfusion distribution in a model of acute lung injury. Acta Anaesthesiol Scand 2011;55:597–606.

[6]. Tusman G, Suarez-Sipmann F, Böhm SH, et al. Monitoring dead space during recruitment and PEEP titration in an experimental model. Intensive Care Med 2006;32:1863–71.

[7]. Rentola R, Hästbacka J, Heinonen E, Rosenberg PH, Häggblom T, Skrifvars MB. Estimation of arterial carbon dioxide based on end-tidal gas pressure and oxygen saturation. J Clin Med 2018;7:290.

[8]. Onodi C, Bühler PK, Thomas J, Schmitz A, Weiss M. Arterial to end-tidal carbon dioxide difference in children undergoing mechanical ventilation of the lungs during general anaesthesia. Anaesthesia 2017;72:1357.

[9]. Bilehjani E, Fakhari S, Yaghoubi A, Eslampoor Y, Atashkhoei S, Mirinajad M. Effect of corrective or palliative procedures on arterial to end-tidal carbon dioxide pressure difference in pediatric cardiac surgery. Afr J Paediatr Surg 2018;73–9.

[10]. Lee JH, Kang PY, Jang YE, Kim EH, Kim JT, Kim HS. A pharmacodynamic model of respiratory rate and end-tidal carbon dioxide values during anesthesia in children. Acta Pharmacol Sin 2019;40:642–7.

[11]. Yang JT, Erickson SL, Killien EY, Mills B, Lele AV, Vavilala MS. Agreement between arterial carbon dioxide levels with end-tidal carbon dioxide levels and associated factors in children hospitalized with traumatic brain injury. JAMA Network Open 2019;2:e199448.

[12]. Guo F, Jin C, Liu S, Congshan Y, Yi Y. Dead space fraction changes during PEEP titration following lung recruitment in patients with ARDS. Respir Care 2012;57:1578–85.

[13]. Gaur P, Harde M, Gujjar P, Deosarkar D, Bhadade R. A study of partial pressure of arterial carbon dioxide and end-tidal carbon dioxide correlation in intraoperative and postoperative period in neurosurgical patients. Asian J Neurosurg 2017;12:475–82.

[14]. Yousuf T, Brinton T, Murtaza G, et al. Establishing a gradient between partial pressure of arterial carbon dioxide and end-tidal carbon dioxide in patients with acute respiratory distress syndrome. J Investig Med 2017;65:338–41.

[15]. Kim YW, Hwang SO, Kang HS, Cha KC. The gradient between arterial and end-tidal carbon dioxide predicts in-hospital mortality in post-cardiac arrest patient. Am J Emerg Med 2019;S0735675718303036.

[16]. Razi E, Moosavi GA, Omidi K, Saebi AK, Razi A. Correlation of end-tidal carbon dioxide with arterial carbon dioxide in mechanically ventilated patients. Arch Trauma Res 2012;1:58–62.

[17]. Fletcher R, Jonson B. Deadspace and the single breath test for carbon dioxide during anaesthesia and artificial ventilation. Effects of tidal volume and frequency of respiration. Br J Anaesth 1984;56:109–19.

[18]. Sivan Y, Eldadah MK, Cheah TE, Newth CJ. Estimation of arterial carbon dioxide by end-tidal and transcutaneous PCO2 measurements in ventilated children. Pediatr Pulmonol 1992;12:153–7.

[19]. Weinger MB, Brimm JE. End-tidal carbon dioxide as a measure of arterial carbon dioxide during intermittent mandatory ventilation. J Clin Monit 1987;3:73–9.

[20]. Whitesell R, Asiddao C, Gollman D, Jablonski J. Relationship between arterial and peak expired carbon dioxide pressure during anesthesia and factors influencing the difference. Anesth Analg 1981;60:508–12.

[21]. Kerr ME, Zempsky J, Sereika S, Orndoff P, Rudy EB. Relationship between arterial carbon dioxide and end-tidal carbon dioxide in mechanically ventilated adults with severe head trauma. Crit Care Med 1996;24:785–90.

[22]. Morley TF, Giaimo J, Maroszan E, et al. Use of capnography for assessment of the adequacy of alveolar ventilation during weaning from mechanical ventilation. Am Rev Respir Dis 1993;148:339–44.

[23]. Tobias JD, Meyer DJ. Noninvasive monitoring of carbon dioxide during respiratory failure in toddlers and infants: end-tidal versus transcutaneous carbon dioxide. Anesth Analg 1997;85:55–8.

[24]. Thrush DN, Mentis SW, Downs JB. Weaning with end-tidal CO2 and pulse oximetry. J Clin Anesth 1991;3:456–60.

[25]. McDonald MJ, Montgomery VL, Cerrito PB, Parrish CJ, Boland KA, Sullivan JE. Comparison of end-tidal CO2 and Paco2 in children receiving mechanical ventilation. Pediatr Crit Care Med 2002;3:244–9.

Keywords:

arterial carbon dioxide tension; end-expiratory carbon dioxide partial pressure; intensive care units; invasive mechanically ventilation

Copyright © 2021 the Author(s). Published by Wolters Kluwer Health, Inc.
Relationship between end-tidal carbon dioxide and arterial... : Medicine (2024)
Top Articles
Latest Posts
Article information

Author: Eusebia Nader

Last Updated:

Views: 6385

Rating: 5 / 5 (60 voted)

Reviews: 83% of readers found this page helpful

Author information

Name: Eusebia Nader

Birthday: 1994-11-11

Address: Apt. 721 977 Ebert Meadows, Jereville, GA 73618-6603

Phone: +2316203969400

Job: International Farming Consultant

Hobby: Reading, Photography, Shooting, Singing, Magic, Kayaking, Mushroom hunting

Introduction: My name is Eusebia Nader, I am a encouraging, brainy, lively, nice, famous, healthy, clever person who loves writing and wants to share my knowledge and understanding with you.